Full-Field Digital Mammography System SternMed – Xenox S200

370

The Xenox S200 is an advanced and re-liable full field digital mammography system with image acquisition stati-on. It is equipped with a transparent protection screen for the operator and is upgradeable with a stereotactic biopsy accessory device. The motori-zed +/-15° rotation of the Iso-centric C-Arm assures accurate, fast and easy workflow.

The main target of the Xenox S200 is to obtain excellent image quality in order to allow sharp visualisation of le-sions and areas of interest minimizing the dose given to the patient. It is suitable both diagnostic as well as for „screening” programs carried out always with utmost accuracy. Excellent imaging technology combined with a modern, ergonomic and winning design improves ef-ficiency and elevates the standard of care.

Manufacturer: Germany


  • Power: 6.6 kVA (0.5 kVA stand-by)
  • Indirect conversion detector technology: a-Si TFT Array + PIN photodiode
  • Optional direct conversion full field flat panel detector FPD technology: amorphous selenium (a-Se)
  • kV range: 20 / 35 kV (20/40 kV optional)
  • kV resolution (manual and automatic modes): 0.5 kV
  • Lowest current time: 1 mAs
  • mAs maximum value: 640 mAs
  • mAs resolution (automatic): 0.1 mAs
  • Exposure time: 02 / 4.7 s (automatically selected depending on the selected mAh)
  • Safety timer: 10 с
  • Auto Collimator: 18 × 24 cm / 24 × 30 cm / 10 × 24 for magnification
  • Automatic filtering (Rh/Ag)
  • Display: 240 × 128 dot graphic LCD
  • Automatic exposure control (AEC)
  • Vertical movements are motorized
  • Motorized rotation
  • ± 15° Rotation for a biopsy procedure
  • Vertical movement range (from the floor): 54 to 145 cm (91 cm stroke)
  • C-arm rotation range: ± 180°
  • C-arm rotation speed in motorized version: 20°/s with acceleration and deceleration ramp
  • Standard compression plate: 18 × 24 / 24 × 30 for normal breasts
  • Optional compression plate: 8 × 24 cm offset sideways for normal breast, 9 × 21 cm straight for magnification, 7.5 cm offset for point contact examination, and 18 × 24 cm offset for two-dimensional biopsy
  • Compression force: adjustable from 70 to 200 N